Tuesday, October 29, 2019
Walking Stick with Heart Attack Detection Essay Example for Free
Walking Stick with Heart Attack Detection Essay 1. INTRODUCTION The National Heart, Lung, and Blood Institute states that ââ¬Å"more than a million persons in the U.S. have a heart attack and about half (515,000) of them die in each year. About one-half of those who die do so within 1 hour of the start of symptoms and before reaching the hospitalâ⬠. A heart attack happens to a person when the blood flow and oxygen supply to heart muscle is blocked, and it is mostly caused by the Coronary Artery Disease (CAD). CAD occurs when the arteries that supply blood to the heart muscle (coronary arteries) become hardened and narrowed. It often causes irregular heart beat or rhythm by blocking blood stream. The National Heart, Lung, and Blood Institute suggest that ââ¬Å"everyone should know the warning signs of a heart attack and how to get emergency helpâ⬠. The symptoms of heart attack can be detected by observing electrocardiogram (ECG) waveform. An ECG is an electrical recording of the heart and is used in the investigation of heart disease. An electrical impulse initiates muscle contraction, which results in heart beating. The spacing between pulses provides a measure of the heartââ¬â¢s rhythm, whereas the height of the pulses is an indicator of pumping strength. By observing the ECG waveform, the heart condition of the patients can be explained by doctors. The ECG Library shows many samples of abnormal ECG waveform, and they are mostly collected from aged people who are more than 55 years old. The senior citizens are more prone to have heart attack than young people. The Walking Stick with Heart Attack Detection is specially designed to help the senior citizens who need walking aids by walking sticks and have the most possibility of heart attack. The walking stick is used as detection unit and as the medium asking for medical help. 1.1 Overview of Design The ECG circuitry unit on the wrist captures abnormal heart beat signal from the patient. The microcontroller on the stick runs a heart attack algorithm. Warning is given out to the person about his heart condition. The Bluetooth emergency calling system calls for medical help at the moment of heart à attack. This project aims to shorten the time between the moment of heart attack and the arrival of medical personal. The warning before the emergency call will give the patient a chance to avoid heart attack. Figure 1.1 Block Diagram of Walking Stick with Heart Attack Detection Two biosensors worn on the userââ¬â¢s wrists send the real ECG signal to the analog ECG circuitry. The amplified and filtered analog output of the circuitry is converted from analog to digital signal and transmitted to the unit on the walking stick. The ECG circuitry unit, the A/D converter, and the transmitter are worn on one of the userââ¬â¢s wrists. The wireless connection between the unit on the wrist and the main unit on the walking stick gives the user more freedom to move by avoiding wire attachment between the wrist and the stick. The receiver on the stick receives the digital ECG signal, and the microcontroller runs a heart attack algorithm to detect possible heart attack symptoms. If any symptom of heart attack is detected, the risk level rises. When the risk level reaches up to the emergency mode, the Bluetooth module activates the userââ¬â¢s mobile phone to call 911 for medical help. Latest mobile phones include GPS function. Therefore, the GPS unit is eliminated from the previous project as the mobile phoneââ¬â¢s GPS can locate the user. 1.2Specifications For the design of the ECG circuitry and the ECG algorithm, the specifications of the previous project ââ¬Å"Wireless Heart Attack Detector with GPSâ⬠were used. The frequency range of ECG signal depends on the activity of individual. The typical range is approximately from 50 Hz to 70 Hz. To cover a wide range of frequencies for all scenarios, the band-pass filter of the ECG circuitry is designed to have a lower cutoff frequency of 0.5 Hz and an upper cutoff frequency of 150 Hz. The analog output of the ECG circuitry must be in the range between -2V to 2V so that the A/D conversion operates properly. The overall amplification inside the ECG circuitry should be at least 3000 but no greater than 5000. The minimum sampling rate of A/D converter should be 400 Hz to capture the detailed ECG waveform that changes in millisecond unit. The transmitted data and the received data must be the same. No noise should be added during the transmission. The distance for reliable transmission should be wide enough so that the movement of user is not limited in the near area of the stick. 40 feet is chosen as the user is assumed to be within this distance during all his activities. All the hardware should operate with low power consumption, and the microcontroller must show the least number of error when it runs the heart attack algorithm. High risk and low risk warnings should be as short as possible. 1.3Performance Benchmarks The total delay from the moment the microcontroller output a heart attack signal to the arrival of medical assistance should be shortened as much as possible. This delay is between the output of heart attack signal and the mobile phone calling. The delay time should be less than 45 second. 1.4Subprojects Figure 1.4 Subproject Flow Chart There are three main subprojects: ECG circuit unit, Analysis Algorithm, and Bluetooth Communication. The first unit is worn on the userââ¬â¢s wrist, and the rest two are installed in the stick. The ECG circuit unit captures ECG waveform from the userââ¬â¢s wrists by electrodes. It then amplifies and filters the ECG signal. After that, the signal is digitized by an A/D converter and is transmitted to the stick. In Analysis Algorithm unit, the receiver first captures the digital ECG signal and feeds it to the microcontroller. The microcontroller runs the algorithm to detect heart attack and to raise the risk level. When a heart attack is confirmed, an emergency signal is sent to the Bluetooth Communication unit. In the Bluetooth Communication unit, the Bluetooth module activates the userââ¬â¢s mobile phone to call 911. Objective: Goals of Design: 1.The wireless heart attack detector captures abnormal heart beat signals. 2.The alert system on the walking stick warns the user to realize his health condition. 3.Bluetooth wireless emergency calling system calls for help at the moment of heart attack via mobile phone. Benefits: Electrocardiogram(ECG) signal transmitted wirelessly from the wrist to the main unit on the stick. This avoids the inconvenience of the attachment of the stick to the wrists. Automatic wireless emergency calling system via Bluetooth module warning mode giving the users a chance to avoid the fatal moment actively. 2.DESIGN PROCEDURE 2.1 Biosensors Since the previous project mentioned about the benefits of the Ag-AgCl ECG electrodes, we decided to use the same electrodes as our biosensors. The benefits of the electrodes include good electrical contact with human skin, low motion artifacts and strong adhesive quality to skin. 2.2 Analog ECG Circuitry Based on the ââ¬Å"Wireless Heart Attack Detector with GPS,â⬠we designed the three components, unity-gain buffers, differential amplifier and band-pass filter. The unity-gain buffers are needed for both wrists as impedance transformers. Although skin impedance is high, the input impedance of op-amps is infinity and the op-amps will be able to catch the bio signals out of the two electrodes. Differential amplifier will take the two bio signals and differentiate them with gain to get the desired ECG waveform. Band-pass filter will make sure that noise of frequencies outside 0.5 Hz and 150 Hz is eliminated. For the Calculation of the ECG circuitry design, the same procedure from the ââ¬Å"Wireless Heart Attack Detector with GPSâ⬠was used and was quoted as below. The differential amplifier gain should not exceed 33 in order to prevent a 300 mV electrode offset potential from causing the system to saturate. The figure shows a diagram of the differential amplifier component of the circuit. Hence the differential amplifier produces an output that is the difference between the inputs multiplied by a gain factor. The gain is chose to be at 15; R7 and R5 are selected to be 15kà ©; R6 and R4 are selected to be 1kà ©. The lower frequency cut-off was desired to be around 0.04Hz, and upper frequency cut-off was desired to be about 150Hz. A gain of around 100 was sought from the band-pass filter. (Overall gain needed to be 1000 to 5000 with the differential amplifier having a gain no greater than 33.) A diagram of the low and high pass portions of the band-pass filter are shown in Figures 2.3 and 2.4, respectively. From eq. (6), it can be seen that the output is amplified by a gain equal to R11/R10. Since the gain is desired to be about 100, eqn. (7) and (8) can be solved. Thus R11 is selected to be 106.2 kà ©, R10 to be 1 kà ©, R8 to be 3.9 Mà ©, C1 to be 1 à ¼F and C2 to be 0.01 à ¼F. The value of R9 is not significant and is selected to be 10 kà ©. A buffer with unity gain is needed for each input as an impedance transformer, because the skin has much higher impedance compared to the input impedance of the differential amplifier. Figure 2.5 shows a diagram of the buffer components. Figure 2.2.4 Buffer Components Diagram Since no current flows into the input terminals and v+ = v-, by applying KCL The gain is determined by R1/R2 = R3/R2. Desiring a gain of unity, R1, R2, R3, and R4 were chosen to be 10 kà © each. 2.3 Data Transmission between Wrist and the Walking Stick Our original design was to transmit the analog ECG waveform directly out of the ECG analog circuitry to the walking stick. We would use the analog signal transmission capability of the HP-3 transceiver. We wanted to make this data transmission wireless from the wrist to the stick. This would avoid the inconvenience of the stick attachment to the wrist. User can go into a car and put his stick at the back seats without detaching any wire between his wrist and the stick. He also does not have to switch the device off. When the stick falls down, it wonââ¬â¢t drag the userââ¬â¢s wrist to the ground. After we browsed through the data sheet of the HP-3 transceiver, we discovered that the analog bandwidth of the transceiver pair is between 50 Hz and 28000 Hz. As discussed in the previous project, normal ECG waveform has frequency range between 50 Hz and 70 Hz. However, to cover all the possible scenarios like sleeping and fast walking, the lower and the upper cutoff frequencies were decided to be 0.5 Hz and 150 Hz. Since the range from 0.5 Hz and 50 Hz is outside the transmission bandwidth of the HP-3 transceiver, we thought about using mixer and oscillator to raise the lowest frequency of analog signal, which is 0.5 Hz, to 60 Hz. This would ensure correct transmission of the analog signal. When we discussed this idea with our TA, we were introduced to the RS232 capability of PIC. Since RS232 signal is digital, we can use the same transceiver to transmit digital signal. We no longer have to worry about the lowest frequency of the analog signal along with the mixer and the oscillator. 2.4 A/D Conversion of Analog ECG Signal to Digital ECG signal Since we are using a PIC for its RS232 feature, we thought it would be natural to use the A/D conversion feature of the PIC. We then had to decide between using an 8-bit conversion or a 10-bit conversion. 10-bit conversion would give higher resolution to the digitized ECG waveform. But, the final decision was determined by the RS232 transmission. According to the PIC-C Compiler Manual, when defining #use RS232, bits sent can only be between 5 and 9. Because 10 are not in this range, we decided to do 8-bit A/D conversion. 2.5 Heart attack detection When the microcontroller on the walking stick has received digital ECG data from the wrists, it will check for heart attack symptoms. We decided to use the heart attack algorithm developed by the previous project. We would like to indicate to the user his heart condition so that he can take proper action like slowing down or taking a rest before heart attack really happens to him. 2.6 Emergency Calling Previous project needed to use a Bluetooth module and a laptop to make an emergency call. Our goal was to eliminate the laptop. Our project will execute emergency calling with just a Bluetooth module. Bluetooth communication is wireless. The user can put his cell phone anywhere he wants as long as it is within the range of the Bluetooth communication. The user does not have to hang his cell phone to the walking stick, making his waking stick heavy and hindering his movement. 3. DESIGN DETAILS 3.1 Analog ECG Circuitry ComponentValue R1, R2, R310 kà © R4, R61 kà © R5, R715 kà © R8100 kà © R93 kà © R101 kà © R11150 kà © C11 à µF C20.01 à µF The band-pass filter has a lower cut-off frequency of 1.59 Hz and an upper cut-off frequency of 106.1 Hz. The gain of band-pass filter is 150, and the overall gain is 150Ãâ"15=2250. 3.2 A/D Conversion and RS232 of PIC16F877 We use PIC16F877 as our microcontroller on the wrist. The connections to the PIC are shown in Appendix 1. Oscillator of 20 MHz is used as clock to the PIC. Analog ECG signal is sent to Pin2 RA0. Pin4 Vref- is connected to 0 V and Pin5 Vref+ is connected to 5 V. The analog signal will be digitized into 8 bits. The digital values will be between 0 and 127. 0 V will correspond to 0, 1 V will correspond to 50, 2 V will correspond to 100 and 2.5 V will correspond to 125. Since the analog ECG waveform will be amplified to only 2 V, the highest digital value will be 100. The reason of not amplifying the signal to 2.5 V is to provide some error of margin in the real world. The sampling will be done with a while loop that runs continuously with the condition always set to ââ¬Ëtrueââ¬â¢. Please refer to the C codes on Appendix 5. The sampling rate of 400 Hz will be implemented by putting a delay of 2500 us in the while loop. After the digital values are obtained through the A/D conversion, the digital values are parallel 8 bits. These parallel 8 bits will be transformed into serial bits of the format of RS232. According to Wikipedia, RS-232 is a standard for serial binary data interchange between a DTE (Data Terminal Equipment) and a DCE(Data Communication Equipment). In our case, the DTE will be the PIC and the DCE will be the HP-3 transmitter. The RS232 format will consist of a ââ¬Ëstartââ¬â¢ bit, eight data bits, least-significant bit first, and a ââ¬Ëstopââ¬â¢ bit. We also make sure the logic voltage level of RS232 will be between 0 V and 5 V as required by the HP-3 transmitter. 3.3 Digital Data Transmission of HP-3 Transceiver The digital ECG values in RS232 form will be fed into a HP-3 transmitter. HP-3 transmitter has 8 parallel selectable channels. To have simple wiring, we selected channel 0 as we only needed to ground the three ââ¬Ëchannel selectââ¬â¢ pins. We are also well aware that the HP-3 transmitter does not encode or packetize the data in any manner. This will assure us that the RS232 data sent from the PIC will be the same RS232 data received at the receiver. The microcontroller at the walking stick does not have to do any decoding. 3.4 Heart Attack Algorithm of the Javelin Stamp Microcontroller The RS232 digital ECG data will be fed from the HP-3 receiver to the Javelin microcontroller. The parameters of the UART of the microcontroller are set up according to those of the PIC. The microcontroller will then get the correct Digital ECG from the PIC. These digital ECG samples will be analyzed with the heart attack algorithm developed by the previous project. Heart attack is preceded by three notable symptoms, which are weak pulse amplitude, irregular heart beat and widened QRS pulse. To check for weak amplitude, a baseline is set up by averaging all the data points. Pulse amplitude is considered as weak if it is below 50% of the baseline. Normal pulse rate is between 35 bpm (beats per minute) and 200 bpm. If the digital ECG data indicates a pulse rate outside of this range, then irregular heart beat is detected. Normal pulse width is less than 36 ms. If the digital ECG data indicates a pulse width more than 36 ms, then QRS pulse has widened 3.5 Low-Risk and High-Risk LEDs of Heart Attack The previous project set up an alert level of threshold of 5 before emergency calling is executed. In our project, we decided to use this alert level to define two conditions, low-risk and high-risk. Low risk is when the alert level is between 4 and 6. High risk is when the alert level is between 7 and 9. These two conditions will be indicated to the user with two LEDs. User can look at the LEDs and be aware of his heart condition to take proper action to avoid the fatal moment. 3.6 EB500 Bluetooth Module and Javelin Stamp Microcontroller When the alert level reaches 10, emergency calling through the userââ¬â¢s cell phone will be executed. In our project, the emergency calling will be done with only the Bluetooth module and the Javelin microcontroller. Laptop will not be used as an intermediate between Bluetooth module and the cell phone. When the Bluetooth module is first powered up, it is in command mode. Communication between the Bluetooth module and the microcontroller will be done in the form of UART. Please refer to the codes in Appendix 6. Microcontroller will send some commands to the Bluetooth module to set up a Bluetooth connection with the userââ¬â¢s cell phone. When the connection is successfully set up, which is indicated by the LED on the Bluetooth module, the module automatically switches to data mode. A successful Bluetooth connection is nothing more than a wireless serial cable. In data mode, everything sent by the Javelin microcontroller will be received by the cell phone. Commands to control a phone are called AT commands. AT commands direct a phone to dial (D), answer (A) and hang up (H). Every AT command starts with ââ¬Å"ATâ⬠(Attention). This is the command line prefix. To make a 911 call, the following command will be sent by the microcontroller to the cell phone, ââ¬Å"ATD911;\râ⬠. The last character, ââ¬Å"\râ⬠is carriage return. The command means ââ¬Å"Attention: Dialing 911â⬠. The number 911 can be replaced by any phone number. Since most cell phones today have the GPS feature, the location of the user can be discovered and faster medical help can be sent to the spot.
Subscribe to:
Post Comments (Atom)
No comments:
Post a Comment
Note: Only a member of this blog may post a comment.